Syringe Size |
10, 20, 30, 50 ml |
Applicable Syringe |
3 brands built-in (BD, B. Braun, Terumo), 2 brands for user-defined |
Volume |
0.1-9999 ml <1000 ml in 0.1 ml increments ≥1000 mL in 1 ml increments |
Flow Rate |
Syringe 10 ml: 0.1-400 ml/h Syringe 20 ml: 0.1-600 ml/h Syringe 30 ml: 0.1-900 ml/h Syringe 50 ml: 0.1-1300 ml/h <100 ml/h in 0.1 ml/h increments ≥100 ml/h in 1 ml/h increments |
Bolus Rate |
Syringe 10 ml: 400 ml/h Syringe 20 ml: 600 ml/h Syringe 30 ml: 900 ml/h Syringe 50 ml: 1300 ml/h |
Accuracy |
±2% (mechanical accuracy ≤1%) |
Infusion Modes |
Flow rate: ml/min, ml/h Time-based Body weight: mg/kg/min, mg/kg/h, ug/kg/min, ug/kg/h |
KVO Rate |
0.1-1 ml/h in 0.1 ml/h increments |
Alarms |
Occlusion, syringe drop off, near empty, end program, low battery, end battery, AC power off, motor malfunction, unattended pump, pressure sensor error, syringe installation error, syringe moved accidently, |
Additional Features |
Syringe identification, real-time shift totals, occlusion pressure release, bolus, mute button, automatic power switching, residue reminder, system memory |
Occlusion Sensitivity |
High: 0.02-0.07 Mpa Medium: 0.05-0.10 Mpa Low: 0.08-0.14 MPa |
Power Supply, AC |
230 V, 50 Hz |
Battery |
9.6±1.6 V, rechargeable |
Battery Life |
5 hours at 5 ml/h |
Working Temperature |
5-40℃ |
Relative Humidity |
20-90% |
Atmospheric Pressure |
860-1060 hpa |
Size |
314*167*140 mm |
Weight |
2.5 kgs |
Safety Classification |
Class Ⅱ, type CF |