Syringe Size
|
10, 20, 30, 50 ml
|
Applicable Syringe
|
3 brands built-in (BD, B. Braun, Terumo), 2 brands for user-defined
|
Volume
|
0.1-9999 ml
<1000 ml in 0.1 ml increments
≥1000 mL in 1 ml increments
|
Flow Rate
|
Syringe 10 ml: 0.1-400 ml/h
Syringe 20 ml: 0.1-600 ml/h
Syringe 30 ml: 0.1-900 ml/h
Syringe 50 ml: 0.1-1300 ml/h
<100 ml/h in 0.1 ml/h increments
≥100 ml/h in 1 ml/h increments
|
Bolus Rate
|
Syringe 10 ml: 400 ml/h
Syringe 20 ml: 600 ml/h
Syringe 30 ml: 900 ml/h
Syringe 50 ml: 1300 ml/h
|
Accuracy
|
±2% (mechanical accuracy ≤1%)
|
Infusion Modes
|
Flow rate: ml/min, ml/h
Time-based
Body weight: mg/kg/min, mg/kg/h, ug/kg/min, ug/kg/h
|
KVO Rate
|
0.1-1 ml/h in 0.1 ml/h increments
|
Alarms
|
Occlusion, syringe drop off, near empty, end program, low battery,
end battery, AC power off, motor malfunction, unattended pump,
pressure sensor error, syringe installation error, syringe moved accidently,
|
Additional Features
|
Syringe identification, real-time shift totals, occlusion pressure release,
bolus, mute button, automatic power switching, residue reminder,
system memory
|
Occlusion Sensitivity
|
High: 0.02-0.07 Mpa
Medium: 0.05-0.10 Mpa
Low: 0.08-0.14 MPa
|
Power Supply, AC
|
230 V, 50 Hz
|
Battery
|
9.6±1.6 V, rechargeable
|
Battery Life
|
5 hours at 5 ml/h
|
Working Temperature
|
5-40℃
|
Relative Humidity
|
20-90%
|
Atmospheric Pressure
|
860-1060 hpa
|
Size
|
314*167*140 mm
|
Weight
|
2.5 kgs
|
Safety Classification
|
Class Ⅱ, type CF
|