PARAMETER
|
SPECIFICATION
|
Drop Detector
|
Outside caliper rotary detector (patent)
Infrared sensor
|
Control Way
|
Drop counting and adjusting the delivery rate automatically
|
Fit IV tubes
|
20d/ml,60d/ml and other, as well as brown tubes
|
Pump Type
|
Peristaltic finger pump
|
Accuracy
|
±2%, when delivery rate ≤25drop/min (drop counting & time control mode)
±3%, when delivery rate >25drop/min (drop counting & time control mode)
±5% (Volume control mode after the correct calibration)
|
Delivery rate range
|
1-1999min (1min step)
|
Volume limit range:
|
0-9999ml (1 ml step)
|
Total Volume Range
|
0-9999.9ml (1 ml step)
|
Purge Rate
|
600.0ml/h
|
KVO Rate
|
KVO is 1ml/h when the delivery rate≤10ml/h, Otherwise KVO is 3ml/h.
|
Occlusion pressure
|
Three level setting LOW, MEDIUM or HIGH
|
Alarms
|
Near empty, Syringe empty, Volume complete (when limit volume is set), Occlusion, Syringe loose, Operation error, Low battery, device failures.
Three levels of alarm volume LOW, MEDIUM or HIGH
|
Bolus function
|
Bolus rate:10-1200ml/h, Bolus volume: maximum of 100ml
|
Power source
|
AC power supply: AC 100 to 240V( +/-10%),50/60Hz;
Built-in Lithium Battery: Rechargeable, 11.1V, 2000mAh.
6 hours of continuous operation
|
Accessories:
|
AC power line, Operational manual
|
Protection Class
|
Class I, Type CF,IPX4
|
Dimension /Net weight
|
130mm*200mm*230mm,2.5kg
|
Add to Enquiry List
|