Features:
1. Universal syringe set: 5ml, 10ml, 20ml, 30ml, 50ml, and 60ml
2. Automatically detection of syringe types and remained solution.
3. Broad range in delivery rate settings.
3. 6 Hours continuous work after fully charged.
4. Three levels of alarm volume: high, medium and low.
5. Three levels of occlusion pressure: high, medium and low.
6. KVO function: After infusion is complete, the pump automatically switches to the smallest flow to prevent creating the thrombus.

 

 

PARAMETER

SPECIFICATION

Applicable syringes

10ml,20ml,30ml,50ml,60ml of any standard

Accuracy

+/- 2% with correct calibration

Rate mode

50ml(60ml):0.1ml/h-1800.0ml/h (0.1ml/h step)
20ml(30ml):0.1ml/h-800.0ml/h (0.1ml/h step)
10ml:0.1ml/h-400.0ml/h (0.1ml/h step)

Time Mode

SOL. VOL:0.1-999.9ml (0.1ml step)
TIME:1-1999min (1min step)

Body Weight Mode:

Volume:0.1-999.9ml(0.1ml step)
DRUG MASS:0.1-999.9mg 0.1mg step)
DOSE:0.01-99.99ug/kg/min(0.01ug/kg/min step)
WEIGHT:0.1-300kg(0.1kgthe least step)

Total Volume Range

0.1-1999.9ml (0.1 ml step)

Purge Rate

50ml(60ml):1800.0ml/h,

20ml(30ml):800.0ml/h,

10ml:400.0ml/h

KVO Rate

1ml/h

Occlusion pressure

Three level setting LOW, MEDIUM or HIGH

Alarms

Near empty, Syringe empty, Volume complete (when limit volume is set), Occlusion, Syringe loose, Operation error, Low battery, device failures. Three levels of alarm volume LOW, MEDIUM or HIGH

Bolus Rate

10ml:10-300ml/h, 20ml(30ml):10-600ml/h, 50ml(60ml):10-1200ml/h

Bolus Volume

Maximum of 25ml

Interface

RS-232, Reserved port

Power source

AC power supply: AC 100 to 240V( +/-10%)50/60Hz;
Built-in Lithium Battery:  Rechargeable, 11.1V, 2000mAh. 6 hours of continuous operation.

Accessories:

AC power line, Operational manual

Protection Class

Class I, Type CF,IPX4

Dimension / Net weight

300mm*240mm*250mm,3.6kg

 

Add to Enquiry List