Overall
Performance
|
Model
|
BIOBASE-CRYSTAL
|
Type
|
discrete, fully automatic
|
Throughput
|
300T/H
|
Analysis Method
|
End-point, Fixed-time,Rate
|
Programming Items
|
programmable for any test items
|
Certification
|
CE, FDA, ISO9001, ISO14001, ISO13485
|
Sample unit
|
Sample Settings
|
60 sample positions, including calibration, quality control and emergency, primary tubes and serum is available
|
Sample Volume
|
2ul ~50ul (0.1ul stepping)
|
Sample Probe
|
Having liquid level induction function, ensuring the accuracy of adding sample
|
Probe cleaning
|
Automatic washing both interior and exterior
|
Analysis Items
|
analyzing maximally 45 items at the same time
|
Reagent unit
|
Reagent Position
|
45 low-temperature refrigerated reagent position
|
Reagent Volume
|
R1:25~300ul, R2:10~150ul (1ul stepping), supporting reagent from different countries
|
Temperature Control
|
continuous 2C~8Ccooling within 24 hours
|
Reagent Probe
|
Having liquid level induction function, ensuring the accuracy of adding reagent. Inventory checking
|
Probe cleaning
|
Automatic washing both interior and exterior
|
Reaction unit
|
Reaction Time
|
10 minutes
|
Temperature Control
|
37C±0.1C
|
Cuvettes
|
120 independent reaction cuvettes (optical path length 6mm)
|
Stirrer
|
after injecting reagent and sample, blending immediately
|
Cleaning
|
automatic water cleaning and drying system
|
System Adjustment
|
System can record the blank value and get more accurate result by deducting blank base value of each cuvette
|
Optical System
|
Light Source
|
12V/30W halogen-tungsten lamp, 2000 hours
|
Receiver
|
8 precision photo receivers
|
Wavelength Range
|
340,405,450,510,546,578,630,700mm
|
Calibration and QC
|
Calibration
|
one-point,multi-point
|
Linear Range
|
0.000~2.500Abs
|
Quality Control
|
Inserting quality control at random.6 quality controls can be reached with one item
|
Software
|
Emergency Function
|
emergency sample can be placed in the sample tray at any time
|
Software
|
Windows 2000 or Windows XP or above, sample and easy to operate
|
Data Processing
|
Having reaction curve detection and data display function in the whole process
|
Others
|
Power Supply
|
AC220V+22V 50Hz+1Hz
|
Dimension
|
1020mm×830mm×1200mm (W*D*H)
|
Weight
|
150kg
|
|
|
|
|
|
|
|